tuitive interface
Minimum maintenance
Main characteristics:
Printer:
Built-in thermal printer. Optional external printer
Parameters:
Intrinsic and extrinsic pathway factors - ATIII - PC - dRvvt - DD
Measurement system:
Measurement principle:
Photometric and turbodensitometric. Coagulometric, chromogenic and turbidimetric methods
Measurement block:
Temperature controlled at 37.4 +/- 0.4ºC (with Peltier system)
2 measurement channels with two wavelengths (405 nm and 750 nm) each available
4 reagent position, one of them with mixing function
18 incubation positions for cuvettes
Automatic start with the addition of the reagent
Test Volume:
Microcuvettes with a capacity of 150-300 µL
Sofware/Hardware:
Open system. Editable and copyable test parameters
15 programmable tests
Automatic calculation of results in seconds, RIN,%, g / L, mg / dL, Ratio, mE, µg / mL
Editable calibration curve (up to 9 points)
One-way Host communication
Multiple languages user interface (DE, EN, ES, FR, PT, RU)