Quality Standards and Classification
CE, ISO13485
SFDA: Class Ⅱb
Anti-electroshock degree:
ClassⅠequipment
(internal power supply)
ECG/Resp: CF
Environment
Operating environment :
Temperature: 0 ~ 40 ℃
Humidity: ≤85%
Altitude: -500 ~ 4600m
Transport and Storage environment :
Temperature: -20 ~ 60 ℃
Humidity: ≤93%
Altitude: -500 ~ 13100m
Power Requirements
AC: 100 ~ 240V, 50Hz/60Hz
DC: Built-in rechargeable battery
Battery: 4000mA 11.1V lithium battery
5min operating after low battery alarm
Date Storage
Trend diagram/table: 7x24h
NIBP review: 400 events
Wave review: 100min
Alarm review: 100 alarm events
Support drug concentration titration analysis
ST Segment
ST Segment Range: -2.0mV ~ +2.0mV
Accuracy: 0.02mV
ECG
5 Leads: RA, LA, LL, RL, V
Lead mode: Ⅰ,Ⅱ, Ⅲ, aVR, aVL, aVF, V
Increase: ×250,×500,×1000,×2000
Sensitivity: > 200 uV (Peak-to-peak value)
Input impedance: > 5 ( megohm )
Bandwidth: Surgery 1 ~ 20 Hz
Monitor 0.5 ~ 40 Hz
Diagnostic 0.05 ~ 130 Hz
CMRR: ≥ 100dB
Polarization Voltage: ± 300mV
Baseline Recovering Time: After defibrillation< 3 seconds
Signal Range: 8 mV(Peak-to-peak value)
Calibrating Signal: 1mV, precision ±5%
RESP
Method: RA-LL impedance
Resp Impedance range: 0.3 ~ 3 Ω
Base Impedance range: 200 Ω ~ 4000 Ω
Bandwidth: 0.1 ~ 2.5Hz
Resp rate: Adult 0 ~ 120BrPM
Neonatal/Pediatric 0 ~ 150BrPM
Resolution: 1BrPM
Precision: ±2 BrPM
Asphyxia Alarm: 10~40 seconds