Technical Specification
Ventilation mode
IPPV, SIMV, SIGH, peep, SIPPV, MANUAL
Ventilator parameter range
Flowmeter: O2 (0.05 ~ 10 L/min)
N2O (0.05 ~ 10 L/min)
Rapid oxygen supply: 35 L/min~75 L/min
Tidal volume (Vt): 0.20 mL~ 1500 mL
Frequence(Freq): 4~99bpm
I:E: 2:1-1:8 Adjustable
PEEP: +0.5KPa --+2KPa
Pressure triggering sensitivity(PTr): -0.3kPa
Flow trigger sensitivity (FTr): 0,.L/min~30 L/min
Pressure control (PC): 5 cmH2O~60 cmH2O
SIGH: 0(off) 1/100~5/100
Apnea Ventilation: OFF, 5s~60 s
Pressure Limit: 20 cmH2O ~100 cmH2O
Monitoring parameter
Frequence(Freq): 0 /min~ 100 /min
Tidal volume (Vt): 0 mL~2000 mL
MV: 0 L/min ~ 100 L/min
Oxygen concentration: 15 %~100 %
Oscillogram
P-T(pressure-time)
V-T(volume-time )
Alarm and protection
The AC power failure alarm: Power failure or no connection
Internal battery backup low voltage alarm: <11.3±0.3 V
No tidal volume: ≤5 mL within 6 s
High oxygen concentration alarm: 19%~100%
Low oxygen concentration alarm: 18%~99%
High Airway pressure alarm: 20 cmH2O ~100 cmH2O
Low Airway pressure alarm: 0 cmH2O~20 cmH2O
High Minute Volume alarm: Adult (5 L/min~ 20 L/min)
Low Minute Volume alarm: Paed (1 L/min~15 L/min, 0~ 10 L/min)
Continuous Pressure alarm: (PEEP+1.5 kPa)over 16s
Suffocation warning: 5s~60s no spontaneous ventilation
The maximum limited pressure: <12.5 kPa
Fan error: Show on screen
Oxygen deficit: Show on screen