Features
7 inch high resolution color touch screen
12-lead simultaneous acquisition and display
ECG Automatic measurement and interpretation function
Complete digital filters, resisting baseline drift, AC and EMG interference
Compact and lightweight design
Support USB flash disk and micro SD card to extend memory
Software upgrade via USB/SD card
Built-in rechargeable Li-ion battery
Specification
Items Specification
Lead Standard 12 leads
Acquisition Mode Simultaneous 12 leads acquisition
Measurement Range ±5mVpp
Input circuit Floating; Protection circuit against Defibrillator effect
Input Impedance ≥50MΩ
Input circuit current 100dB
Patient leak current <10μA(220V-240V)
Frequency Response 0.05Hz~150Hz(-3dB)
Sensitivity 2.5mm/mV, 5 mm/mV, 10 mm/mV, 20 mm/mV(error: ±5%)
Anti-baseline Drift Automatic
Time constant ≥3.2s
Noise level <15μVp-p
Paper speed 12.5 mm/s, 25 mm/s, 50 mm/s
Recording mode Thermal printing system
Paper specification Roll 110mmx20M
Safety Standard IEC I/CF
Sample Rate Normal: 1000sps/channel
Power Supply AC: 100~240V, 50/60Hz, 30VA~100VA
DC: 14.8V/2200mAh, built-in lithium battery