The power of generators 32KW, DR generating filming, the sliding column x-ray system comes complete with a four way floating top table with integral under surface cassette tray.
The counter balanced adjustable height tube column, supporting the dual focus rotating anode Toshiba tube, not only enables the focal film distance to be adjusted to any height above the table top, but slides in floor mounted runners beyond the full length of the table making standing lateral images easier than ever.
Generator
Power voltage
1-phase 220VAC ±10%,50Hz,16A;3-phase 380VAC±10%
Maximum Working Frequency
≥30KHz
Nominal Electric Power
32kw(@100kv,200mA,100ms)
KV range
40-125KV
mA range
10-320mA
mAs range
0.1-400mAs
Photo form time
3ms-6300ms
X-ray tube
Working voltage
40KV-125KV
Maximum power (0.1 S)
Small focus 20 kw, Large focus 40 kW
Nominal value
Large focus 1.2,Small focus 0.6
Maximum heating filament characteristics
5.3A
Anode speed (50HZ)
2800rpm
Anode angle
12°
Inherent filtering
1.0mmAL/75kV
Anode Heat Content
110KJ(150KHU)
Tube component heat capacity
900KJ(1300KHU
Flat Panel Detectors
Detector Technology
a-Si
Scintillator
CsI
Image Size
43×43 cm
Pixel Matrix
3072×3072 (17×17 in)
Pixel Pitch
140 μm
A/D Conversion
16 bit
Minimum Detectable Dose
14 nGy
Maximum Linear Dose (RQA5)
110 μGy
Spatial Resolution
3.6 LP/mm
Data Interface
GigE /802.11ac
Power Dissipation
20 W
Adapter Input
AC 100-240V,50-60Hz
Dimensions
46×46×1.5 cm