Efficient filtration and kV settings for the lowest dose possible Wireless flat panel detector Precise and clear imaging Impressively flexible range of movement Light and narrow (1000mm wide) Adaptable to various terrains High frequency generator (32kW) Long battery life (can last up to 30km) Superb anode heat capacity (300KHU) Intuitive user interface 19” LCD display Designed for an optimized workflow With an advanced image stitching algorithm
Flat Panel Detector
(1) Type of the detector: TFT monolithic amorphous silicon
(2) Panel size of the detector: 430mm*430mm
(3) Efficient imaging size of the detector: 17*17 inch
(4) Acquisition pixel matrix of the detector: ≥3072×3072
(5) Pixel pitch of the detector: ≤146 um
(6) Spatial resolution of the detector: 3.4LP/MM
(7) Pixel gray scale: 16 Bit
(8) Time of image acquisition and transmission: 3s
2. High frequency and high voltage generator
(1) Max. Output power: 32kw
(2) Input power supply: 220 VAC single phase
(3) Output voltage: 40~150kV
(4) Tube current range: 10~320mA
(5) Exposure time range: 1ms-10000ms
(6) mAs range: 0.1mAs~250mAs
(7) Support AEC,APR automatic exposure
(8) Diagnostic self-test and display
3. X-ray bulb tube assembly
(1) Focus: 0.6mm/1.5mm
(2) Nominal electric power: 21kw/46kw (50Hz)
(3) Max. KV: 150kV
(4) Anode target angle: 12°
4. Column
(1) Rotation angle range of the column is -145°~180°, deviation±2°;
(2) Up and down rotation angle of the bulb tube and the beam limiting device on horizontal axis: -90°~45°, deviation±2°;
(3) Vertical moving stroke of the X-ray tube assembly focus: 95cm, deviation±2cm;